Application Range
Standard Measuring Range: 30 to 400 ppm
Number of Strokes n: 6
Time for Measurement: approx. 2 min
Standard Deviation: ± 10 to 15%
Color Change: white -› brown
Ambient Operating Conditions
Temperature: 10 to 30°C
Absolute Humidity: 5 to 12 mg H2O / L
Reaction Principle
C6H5 - C2H5 + I2O5 -> I2
Cross Sensitivity
Many petroleum hydrocarbons and aromatics are indicated, but with different sensitivities. It is impossible to differentiate between them.